BINDER LITMK240-400V Series LIT MK - Battery Test Chambers, With Safety Equipment For Rapid Temperature Changes, 9020-0402
Quick Overview
Average cooling down time according to IEC 60068-3-5 [K/min]: 3.5 Average heating-up rate according to IEC 60068-3-5 [K/min]: 5 Cooling down time from 110 °C to -40 °C [min]: 160 Max. heat compensation at 25 °C [W]: 3000 Temperature fluctuation depending on setpoint [± K]: 0.1…0.5 Temperature range: -40…110 Temperature variation depending on setpoint [± K]: 0.1…1.2
Details
Benefits
Safety thanks to extensive protective measures Homogeneous climate conditions thanks to APT.line™ preheating chamber Extensive programming and data collection Excellent overview of the test object thanks to large heated viewing window
Important characteristics
Temperature range: -40 °C to +110 °C Class 2 independent temperature safety device when temperature is set to +120 °C Reversible stainless steel pressure relief flap installed in the middle on top of the unit Enhanced door-locking mechanism with reinforced brackets Monitoring of CO, H₂, O₂, and temperature in testing area atmosphere Inertization equipment CO₂ fire suppression APT.line™ preheating chamber technology Programmable condensation protection for samples Heated viewing window with LED interior lighting BINDER Multi Management Software APT-COM™ basic edition Intuitive touchscreen controller with time-segment and real-time programming Internal data logger, measured values can be read out in open format via USB 2 access ports with silicone plugs, 80 mm, left and right Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm Computer interface: Ethernet 1 stainless steel rack, can be freely positioned CFC-free refrigerant R-452A Cooling with compressor cooling unit TÜV SÜD confirms that the LIT MK 240 and LIT MK 720 chambers are suitable for EUCAR level 6
Optional equipment
Electromechanical door-locking mechanism Additional access ports available in various sizes and locations Multi Management Software APT-COM for managing, recording, and documenting unit parameters Services – wide range of services to ensure that the unit functions correctly
Technical specifications
Performance Data Temperature:
Average cooling down time according to IEC 60068-3-5 [K/min]: 3.5 Average heating-up rate according to IEC 60068-3-5 [K/min]: 5 Cooling down time from 110 °C to -40 °C [min]: 160 Max. heat compensation at 25 °C [W]: 3000 Temperature fluctuation depending on setpoint [± K]: 0.1…0.5 Temperature range: -40…110 Temperature variation depending on setpoint [± K]: 0.1…1.2
Electrical data
Nominal power [kW]: 5.6 Phase (Nominal voltage): 3~ Power frequency [Hz]: 50 Rated Voltage [V]: 400 Unit fuse [A]: 16
Doors
Unit doors: 1
Housing dimensions not incl. fittings and connections